Finesse +

Finesse®+ Electrosurgical Generators are the only systems with integrated smoke evacuation. Their unique electronic design improves results for office-based procedures of gynecologists and other medical specialties.

Product Description
Models
Specifications
Specialized Gynecology Brochure
Finesse® Brochure


The Finesse+ and Finesse II+ Electrosurgical Generator and Smoke Evacuation Systems have been re-designed to meet the highest performance and safety standards currently required for electrosurgery.

Finesse +

Finesse II +

Finesse® + LEEP Surgery Electrosurgical Generator with Integrated Smoke Evacuator.
Finesse® II+ LEEP Surgery Electrosurgical Generator with Integrated Smoke Evacuator.

CONTROLLED OUTPUT CIRCUITRY+

UTMD’s electrosurgical experience and research into tissue effects during loop electrosurgery have resulted in an improvement to system design. Finesse+ and Finesse II+ incorporate Controlled Output Circuitry+ to produce the best tissue specimen for conclusive histopathology. Controlled Output Circuitry+ is UTMD’s advancement of “intelligent cut” circuitry that maintains the output within a prescribed cutting range by monitoring and continuously adjusting the output to produce a specimen with minimal thermal damage at the margins. This eliminates any need to adjust the output setting when changing loop sizes.

Controlled Output Circuitry+ is a three-tier output delivery and monitoring scheme:

Tier 1: A microprocessor and specialized electronics continuously monitor the output, adjusting it to remain at the ideal level for smooth, char-free cutting.

Tier 2: The microprocessor compares the output to mathematically-defined reference curves, and further adjusts the output as necessary to ensure that safe output levels are maintained.

Tier 3: In the event that output cannot be adjusted to satisfy the reference curves, output is disabled and an error is indicated.

INTEGRATED SMOKE EVACUATION

The Finesse+ and Finesse II+ Systems utilize a design that integrates the electrosurgical generator and smoke evacuation system into a single compact unit. This allows placement in operating areas with limited space, especially offices. It also allows simultaneous “single switch” activation of both modules by either the handswitch control pen or footswitch.

Finesse+ and Finesse II+ use a three-stage filtration system to evacuate and filter the smoke plume produced during electrosurgery. The filtration system includes an activated charcoal filter which adsorbs odorous gases, and two high-efficiency particulate filters which remove solid particles and aerosols. The three filters provide a minimum efficiency of 99.999% for 0.1 micron particles.

DISPERSIVE PAD CONTACT QUALITY MONITORING (FinCQM™)

UTMD’s FinCQM circuit design adjusts to skin type variations and detects partial pad detachment before a pad site burn can occur. Output is automatically disabled and an error is indicated with separation of approximately 30% of the pad surface.

PATIENT AND USER SAFETY

The Finesse+ and Finesse II+ Systems meet global standards for patient lead isolation. This provides protection for both patient and clinician by reducing the possibility of creating an alternate current path that could result in a burn.

ENHANCED LOGIC INTEGRATION

Digitally-driven waveform shapes and a majority of the logic functions are hard-coded into a microprocessor-linked complex programmable logic device (CPLD). Component usage is significantly minimized to ensure reliable operation of the Finesse+ and Finesse II+ systems.

The Finesse + Electrosurgical Generator and Smoke Evacuation System

The Finesse + Electrosurgical Generator and Smoke Evacuation System

Electrosurgical Generators
Model Number Description Unit of Measure Image
FIN-110 Finesse+ Electrosurgical Generator
(115 VAC)
1/EA Finesse® + LEEP Surgery Electrosurgical Generator. Models ESU-110 & ESU-220
FIN-220 Finesse+ Electrosurgical Generator
(230 VAC) *
1/EA
FIN2-110 Finesse II+ Electrosurgical Generator
(115 VAC)
1/EA Finesse® II+ LEEP Surgery Electrosurgical Generator. Models FIN2-110 & FIN2-220
FIN2-220 Finesse II+ Electrosurgical Generator
(230 VAC) *
1/EA
 

* Product is CE Marked

Finesse+ (FIN-110 / FIN-220)
Dimensions:
Width 14.0" (35.6 cm)
Depth 14.7" (37.3 cm)
Height 7.3" (18.5 cm)
Weight 24 lb. (11 kg)
Electrical Supply:
Voltage 115 volt / 230 volt
Frequency 50 Hz / 60 Hz
Current 5.65 Amps / 3.75 Amps
Low Frequency Leakage <100 microamps
Electrical Output:
Frequency 450 kHz
Cut/Blend Power 6 - 99 Watts @ 500 ohm load
Cut Mode  
Cut Continuous Sinusoid
Blend 1 Interrupted Sinusoid: 62.5% Duty Cycle
Blend 2 Interrupted Sinusoid: 50% Duty Cycle
Blend 3 Interrupted Sinusoid: 37.5% Duty Cycle
Coag Power 6 - 75 Watts @ 500 ohm load
Coag Voltage 2,400 Volts zero-to-peak max (open circuit)
Dispersive Pads:
Compatible Types Auto-detects and displays pad type: Dual (CQM) or Standard
CQM Circuit Initial threshold detect, with threshold auto-adjust with improved contact. 10-130 ohms operating range
Error Indicators:
CQM System Unacceptable pad peel
  Pad contact out of range
Dispersive Pad Status Pad not connected to system
  Pad type mismatch
Output Monitor Hazardous output power limit
  Output current limiting circuit
Cross-Key Simultaneous Cut/Coag activation
Mode Change Mode change during activation
Power Adjust Control lockout during activation
Activation:
Type Handswitch, Footswitch (with optional two-pedal footswitch)
Smoke Evacuation:
Flow Rate  
Normal >70 liters/min. (2.5 CFM)
High >100 liters/min. (3.5 CFM)
Efficiency >99.999% @ 0.1 microns
Standards Compliance: IEC 60601-1 (3rd ed) + IEC 60601-2-2 (5th ed) (electromedical safety)
IEC 60601-1-2 (electromagnetic compatibility)
 
Finesse II+ (FIN2-110/FIN2-220)
Dimensions:
Width 14.0" (35.6 cm)
Depth 14.7" (37.3 cm)
Height 7.3" (18.5 cm)
Weight 24 lb. (11 kg)
Electrical Supply:
Voltage 115 volt / 230 volt
Frequency 50 Hz / 60 Hz
Current 5.65 Amps / 3.75 Amps
Low Frequency Leakage <100 microamps
Electrical Output:
Frequency 450 kHz
Cut Power 65 Watts @ 500 ohm load
Cut Mode Blended Cut, Interrupted Sinusoid: 62.5% Duty Cycle
Coag Power 60 Watts @ 500 ohm load
Coag Voltage 2,180 Volts zero-to-peak max (open circuit)
Dispersive Pads:
Compatible Types Auto-detects and displays pad type: Dual (CQM) or Standard
CQM Circuit Initial threshold detect, with threshold auto-adjust with improved contact. 10-130 ohms operating range
Error Indicators:
CQM System Unacceptable pad peel
  Pad contact out of range
Dispersive Pad Status Pad not connected to system
  Pad type mismatch
Output Monitor Hazardous output power limit
  Output current limiting circuit
Cross-Key Simultaneous Cut/Coag activation
Activation:
Type Handswitch, Footswitch (with optional two-pedal footswitch)
Smoke Evacuation:
Flow Rate >80 liters/min. (2.8 CFM)
Efficiency >99.999% @ 0.1 microns
Standards Compliance: IEC 60601-1 (3rd ed) + IEC 60601-2-2 (5th ed) (electromedical safety)
IEC 60601-1-2 (electromagnetic compatibility)
 

 